Based on your previous selection, you have been forwarded to the website for International.

XL ROS1-GOPC BA

Break Apart Probe

Order Number
D-6029-100-OG
Package Size
100 µl (10 Tests)
Labels
  
Chromosome
6
Regulatory Status
IVDD

IVDR Certification

MetaSystems Probes has already certified a large part of its portfolio, according to IVDR. For organizational reasons, we currently provide only the IVDD product.

Discover all IVDR-certified products

Product Description

XL ROS1-GOPC BA

XL ROS1-GOPC BA consists of a green-labeled probe hybridizing proximal to the ROS1 gene region at 6q22.1 and an orange-labeled probe hybridizing distal to the ROS1 gene region at 6q22.1.

Probe maps are created in accordance with the intended purpose of the product. Solid colored bars do not necessarily indicate that the probe fully covers the indicated genomic region. Therefore, caution is advised when interpreting results generated through off-label use. Probe map details based on UCSC Genome Browser GRCh37/hg19. Map components not to scale. Further information is available on request.

Need assistance or a quote?

Clinical Details

Clinical Applications

  • Solid Tumors (Solid Tumors)
Need assistance or a quote?

Images

XL ROS1-GOPC BA

XL ROS1-GOPC BA hybridized to the U-138 cell line. Several interphase nuclei show the expected fusion pattern, while one cell to the left is showing a homozygous deletion of the orange signals.

Need assistance or a quote?

Expected Patterns

Expected Pattern 1

Normal Cell:
Two green-orange colocalization/fusion signals (2GO).

Expected Pattern 2

Aberrant Cell (typical results):
One green-orange colocalization/fusion signal (1GO), one separate green (1G) and orange (1O) signal each resulting from a chromosome break in the respective locus.

Expected Pattern 3

Aberrant Cell (typical results):
One green-orange (1GO) colocalization/fusion signal and one green (1G) signal resulting from the loss of one orange signal.

Need assistance or a quote?

Literature

  • Charest et al (2003) Genes Chrom Cancer 37:58-71
  • Takeuchi et al (2011) Nat Med 18:378-381
  • Bergethon et al (2012) J Clin Oncol 30:863-870

Downloads

News


IVDR Certification

MetaSystems Probes has received IVDR certification for our initial 26 fluorescence in situ hybridization (FISH) probes from the notified body, BSI. Achieving this milestone was not without its challenges, and we are delighted to have accomplished IVDR certification for this probe set at an early stage.

IVDR Certification